AC current

 

Range Resolution Accuracy Overload protection
999.9mA 0.1mA ±(1.2%+3) Fused
9.999A 0.001A
  • Theidentified minimum current is about 2mA.
  • Virtualvalue of sine wave,AC frequency response is 40-400Hz. Accuracy   scope:5%~100%range(manual   mode).
  • ACcrestfactor or AC crest factor of non-sinusoidal wave add an error based on bellows: 1)lf crest factor is 1~2,add 3%.

2)lf crest factor is 2~2.5,add 5%. 3)lf crest factor is 2.5-3,add 7%.

Attention:Please make sure the probe is away from the meterbeforeopening thebottomcover. 1.If   the    meteris    not    in    use,please    turnit    of    to   avoidbattery   consumption.

2.Maintenance

a.The meter mustbe maintained or repaired by professionals or specified repair centers.

b.Clean the cover by dry cloth on a regular basis.Do not use any grinders or solvents.

3.Replace batteries or fuse(Figure 4a and Figure 4b).The meter uses 2 pieces of AAA1.5V batteries,please install or replace batteries according to the sequence bellow:

a.When the meter is turned off,please make su ure tlhe measured probe is away from the meter.

b.Make sure the meter backside is upward,unscrewing,opening the cover,taking out batteries, new batteries must be installed according to the polarity.

c.Installing new batteries,cover batteries and screw the cover.

d.Unscrew the rear cover to replace a new fuse(10A/600V ceramic fuse,φ 6×25mm).

Figure 4a                                                  Figure  4b